TensION

Test Method

Hold the TensION at a distance from the ionising points / emitters. If a high voltage is present, a red indicator in the plastic tip will illuminate. To get an indication of proper functioning of the Simco-Ion equipment all the emitters should be checked.
Minimum distance from the TensION to the ionising emitters or charging points must be obtained. Refer to the table below:

Test Method

Hold the TensION at a distance from the ionising points / emitters. If a high voltage is present, a red indicator in the plastic tip will illuminate. To get an indication of proper functioning of the Simco-Ion equipment all the emitters should be checked.
Minimum distance from the TensION to the ionising emitters or charging points must be obtained. Refer to the table below:

Operating Voltage Minimum distance
3,3 kV 5 mm
4 kV 6mm
5 kV 10 mm
7 kV 13 mm
10 kV 20 mm
30 kV 80 mm

Technical specifications

Power 2 size AAA batteries (included)
Measuring distance contactless (see table above)
Weight 40 grams
Material Radilon-S HS(a), BE 170MO IP64
Dimensions Length 148 mm, diameter maximum 17 mm
Ambient temperature -10°C - +55°C